IV Administration Set - With Universal Spike and Spin-Lock Connector
IV Administration Set - With Universal Spike and Spin-Lock Connector
SKU:
14 in stock
We have 3 Options: 79in, 84in, and 85in
Option#1: B-Series Set, v1402
Length: 79in, 201cm
REF: V1402 | GTIN: 04046964188183
Instructions on Package:
Use Aseptic Technique
Replace set as per CDC Guidelines
1. Close roller clamp. Fully spike container. Hang fluid container. Gently squeeze and release drip chamber until approximately half full.
NOTE: Vent cap should remain in the closed position unless infusing from a solid stopper glass bottle.
2. Prime set by slowing opening roller clamp and allow fluid to fill tubing. Close roller clamp. Ensure air is expelled. Repeat prime if necessary. Attach infusion Device.
3. Slowly open roller clamp and adjust for desired flow rate.
-
Option#2: v1421-20, 84in
Length: 84in, 213cm
REF: V1421-20 | GTIN: 04046964188244
Instructions on Package:
Use Aseptic Technique
To access Injection site: Swab injection site vigorously for 15 seconds and allow to air dry prior to use. Add compatible medication by injecting through injection site.
Replace set as per CDC Guidelines.
1. Close roller clamp. Fully spike container. Hang fluid container. Gently squeeze and release drip chamber until approximately half full.
NOTE: Vent cap should remain in the closed position unless infusing from a solid stopper glass bottle.
2. Prime set by slowing opening roller clamp and allow fluid to fill tubing. Close roller clamp. Ensure air is expelled. Repeat prime if necessary. Attach infusion Device.
3. Slowly open roller clamp and adjust for desired flow rate.
-
Option#3: 352601, v1501, 85in
Length: 85in, 216cm
REF: v1501, 352601 | GTIN: 04046964181795
Instructions on Package:
Use Aseptic Technique
Replace set as per CDC Guidelines.
1. Close roller clamp. Fully spike container. Hang fluid container. Gently squeeze and release drip chamber until approximately 1/3 full.
NOTE: Vent cap should remain in the closed position unless infusing from a solid stopper glass bottle.
2. Prime set by slowing opening roller clamp and allow fluid to fill tubing. Close roller clamp. Ensure air is expelled. Repeat prime if necessary.
-
For Gravity Administration
3. Attach to infusion device.
4. Slowly open roller clamp and adjust for desired flow rate.
For Use with VISTA basic pump
Cautions:
Do not use same tubing section in pump for more than 24 hours.
Unrestricted fluid flow may occur if the set is not properly installed in the pump.
Set does not contain set based free flow protection.
Refer to VISTA basic Instructions for use.
1. Proceed with steps 1&2 in prior section.
2. Open the door of the pump. Push and hold open the clamp at the top of the fluid path. Insert tubing in clamp and release clamp.
3. Thread the tubing down the fluid path making sure the tubing is pressed into the Air-in-Line Detector at the bottom of the fluid path.
4. Close the door of the pump.
5. Open all clamps on the tubing. Check that no drops are falling into the drip chamber. Attach infusion device.
6. Proceed with pump operating device.